ELISA Protocol

Gain Data - ELISA analysis tool, web based free service
Do you have any issues with your ELISA? – contact us at Info@2BScientific.com!

View PDF Version

1. Introduction

ELISA (enzyme-linked immunosorbent assay) is a plate-based assay for the detection and quantification of peptides, proteins, antibodies and hormones. The ELISA technique is based on an antigen immobilised to a solid surface, which is complexed with an antibody that is linked to an enzyme.

The enzyme activity upon incubation with a substrate and production of a measurable product is used for the detection and quantification of the factor of interest. This detection method relies on the highly specific antibody-antigen interaction. ELISAs are typically carried out in 96-well (or 384-well) polystyrene plates, which will passively bind antibodies and proteins. The binding and immobilization of reagents of interest is easy to design and perform. The subsequent separation of specifically bound from unbound factors makes ELISA assays a powerful tool for the analysis of relevant analytes.

2. General ELISA Procedure

Typically, an ELISA begins with a coating step, in which the first layer, consisting of a target antigen or antibody is adsorbed onto a 96-well polystyrene plate, unless you are using a kit with a plate that is pre-coated with antibody. Following, all unbound sites are coated with blocking agent. After a series of washes, the plate is incubated with enzyme-conjugated antibody. Another series of washes removes all unbound antibody.

The addition of a substrate causes the production of a calorimetric signal, which is finally detected. Because the separation of factors of interest from a heterogeneous sample relies on specific surface binding, several washes are repeated in each ELISA step to remove unbound material. During this process, it is crucial that excess liquid is removed in order to prevent the dilution of the solutions added in the next step. To ensure uniformity, specialised plate washers can be utilised.

An ELISA assay can include multiple intervening steps, particularly when measuring protein concentration in heterogeneous samples such as blood. The detection of multiple antibodies for signal amplification at the same time represents the most complex and varying step in the overall process.

3. ELISA Types

LISAs can be performed as direct, indirect, sandwich or competitive assays. The immobilization of the antigen, the key step of the ELISA method, can be achieved by direct adsorption to the assay plate or indirectly via capture antibody that has been attached to the plate. Further, the antigen is then either detected directly with an enzyme-labeled primary antibody or indirectly by an enzyme-labeled secondary antibody.

The antibodies used for detection are commonly coupled with alkaline phosphatase (AP) or horseradish peroxidase (HRP). Dependent on the required assay sensitivity and the signal-detection method (spectrophotometer, fluorometer or luminometer), a variety of substrates are available for performing the ELISA with an HRP or AP conjugate. Hence, when designing an ELISA assay, the different ELISA strategies, specifically for the detection step, have to be differentiated and adjusted to the requirements of the nature of the sample and the biological question.

Direct ELISA

In a direct ELISA assay the antigen of interest is immobilised to the plate and detected with an enzyme-coupled antibody, while a blocking agent is used to occupy unspecific binding sites. This approach is used for the detection and quantification of non-commercial proteins.

+ Quick protocol due to the use of only one antibody.
+ No possibility of cross-reacting secondary antibody.
- Immunoreactivity of the primary antibody could be adversely affected by labelling with enzymes or tags.
- Time-consuming and expensive labelling of primary antibodies for each specific ELISA system.
- Constraint to the labelling method of the primary antibody label from experiment to experiment.
- Minimal signal amplification.

Indirect ELISA

For indirect detection, the antigen coated to a multi-well plate is detected in two stages or layers. First an unlabeled primary antibody, which is specific for the antigen, is applied. Next, an enzyme-labeled secondary antibody is bound to the first antibody. The secondary antibody is usually an anti-species antibody and is often polyclonal. The indirect assay, the most popular format for ELISA, has the following advantages and disadvantages:

+ A wide range of labeled secondary antibodies is commercially available.
+ Versatile approach, as many primary antibodies can be generated in one species and the secondary antibody, with the same label can be used for detection.
+ Highest possible immunoreactivity of the primary antibody is retained because it is not labeled.
+ Increased sensitivity as the primary antibody contains several epitopes that can be bound by the labeled secondary antibody, enhancing the signal.
- The secondary antibody might cross-react and cause a nonspecific signal.
- An additional incubation step is required for the procedure.

Sandwich ELISA

In a sandwich ELISA two antibodies, specific for different, non-overlapping epitopes are used: firstly, the plate is coated with a capture antibody. In the next step the sample is added and bound by the capture antibody. The signal identification is achieved by adding the secondary detection antibody.

+ High specificity due to the method of specifically capturing and detecting the antigen.
+ Suitable for complex (or heterogeneous/impure) samples: the antigen does not require purification prior to measurement.
+ High flexibility and sensitivity as both direct and indirect detection methods can be used.

Competitive ELISA

The method of the competitive ELISA is based on the competitive reaction between the sample antigen and the antigen bound to the microtiter plate, labeled with the primary antibody. Initially, the primary antibody is incubated with the sample antigen and the formed antibody-antigen complexes are added to the antigen coated microtiter plates. Following an incubation step, any unbound antibody is washed off.

The more antigen present in the sample, the more primary antibody will be bound to the sample antigen. Conversely, there will be a smaller amount of primary antibody available to bind to the antigen coated on the well, causing a signal reduction.

The main advantage of this ELISA method arises from its high sensitivity to compositional differences in complex antigen mixtures, even when the specific detecting antibody is present in relatively small amounts.


 IndirectDirectSandwichCompetitive
Capture Antibody CoatingXXX
Antigen CoatingX
Blocking
Sample (Antigen) IncubationXX
Primary Antibody Incubation
Secondary Antibody IncubationX
Substrate Preparation
Signal Detection
Data Analysis

4. ELISA results 
The ELISA assay produces 3 types of data output:
1) Qualitative: ELISA can give some information about the presence of a particular antigen in a sample, as compared to a blank well containing no antigen or an unrelated control antigen.

2) Quantitative: the concentration of antigen in the sample is inferred from a comparison to a standard curve (serial dilution of an antigen of known concentration).

3) Semi-Quantitative: ELISAs can be used to compare the relative levels of antigen in assay samples, since the intensity of signal will vary directly with antigen concentration.

ELISA data is typically graphed with optical density against the log concentration to produce a sigmoidal curve. Known concentrations of antigen are used to generate a standard curve which is further used to measure the concentration of unknown samples by comparison to the linear portion of the standard curve.

Effectively, it is the relatively long linear region of the curve that makes the ELISA results accurate and reproducible. The unknown concentration can be determined directly on the graph or with curve fitting software, which is typically found on ELISA plate readers.

5. Sample Preparation

The procedure below provides a general guidance for the preparation of commonly tested samples for use in ELISA assays.
Generally:
- Protein extract concentration should be at 1-2 mg/mL.
- Cell and tissue extracts are diluted 1:1 with binding buffer.
- Samples are centrifuged at 10,000 rpm for 5 min at 4°C before use, to remove any precipitate.

• Cell Culture Supernatants
Centrifuge cell culture media at 1,500 rpm for 10 min at 4°C, aliquot and assay supernatant immediately. Hold at -80°C and avoiding freeze thawing.

 Cell Extracts
Place tissue culture plates on ice. Remove the media and gently wash cells once with chilled (10 mM Na2HPO4 and 1.8 mM NaH2PO4 in deionized water with 0.2% Tween 20; pH Adjusted to 7.4). Remove the PBS and add 0.5 ml extraction buffer per 100 mm plate. Tilt the plate and scrape the cells into a pre-chilled tube. Vortex briefly and incubate on ice for 15-30 min. Centrifuge at 13,000 rpm for 10 min at 4°C to pellet the insoluble content. Aliquot the supernatant in chilled tubes on ice, store at -80°C and avoid freeze thawing.

• Conditioned Media
Plate the cells in complete growth medium (with serum) until the desired level of confluence is achieved. Remove the growth media and gently wash cells twice with 2- 3 ml of warm PBS. Remove the PBS, gently add serum-free growth medium and incubate for 1-2 days. Transfer the medium into a centrifuge tube and centrifuge at 1,500 rpm for 10 min at 4°C. Aliquot the supernatant, store samples at -80°C and avoid freeze thawing.

• Tissue Extract
Homogenize tissue on ice in ice-cold buffer, containing protease inhibitors. Place the tissue in micro-centrifuge tubes and snap-freeze in liquid nitrogen. Store samples at -80°C or keep on ice for immediate homogenization. Add 300 μL of extraction buffer per 5 mg of tissue and homogenize.

 Extraction buffer
00 mM Tris, pH 7.4
150 mM NaCL
1 mM EGTA
1 mM EDTA
1% Triton X-100 0.5%
0.5% sodium deoxycholate (SDS)

Extraction buffer can be prepared in advance and stored at 4°C. Add phosphatase inhibitor (according to manufacturers guidelines), protease inhibitor (according to manufacturers guidelines) and PMSF to 1 mM before use.

Rinse the blade of the homogenizer twice with 300 μL extraction buffer. Place the sample on a shaker at 4°C for 2 hours. Centrifuge the sample for 20 min at 13,000 rpm at 4°C. Aliquot the supernatant into pre-chilled tubes on ice. Keep the samples at -80°C and avoid freeze-thawing
Note: Lysis buffer volume must be determined according to the amount of tissue present. Typical concentration of final protein extract is at least 1 mg/mL.

6. Recommended Protocols

Reagent Preparation:

Standard Solutions (use within 2 hours) For the standard concentration of 10,000 pg/mL add 1 mL of sample diluent buffer into one tube of standard (10 ng per tube) and mix thoroughly.
Note: Store this solution at 4°C for up to 12 hours (or -20°C for 48 hours) and avoid freeze-thawing.

For the standard concentration of 5,000 pg/mL add 0.3 mL of 10,000 pg/mL to 0.3 mL of sample diluent buffer and mix thoroughly.

For the standard concentration of 2,500 pg/mL add 0.3 mL of 5,000 pg/mL to 0.3 mL of sample diluent buffer and mix thoroughly.

Perform similar dilutions to obtain the standard solutions with the concentrations of 1,250, 625, 312, 156 and 78 pg/mL.

Add 100 μL of each of the diluted standard solutions to the appropriate empty wells. Repeat in duplicate or triplicate for accuracy.

Biotinylated Antibody
Calculate the total volume needed for the assay by multiplying 0.1 mL/well and the number of wells required. Add 2-3 extra wells to the calculated number of wells to account for possible pipetting errors. Generate the required volume of diluted antibody by performing a 1:100 dilution (For each 1 μL concentrated antibody, add 99 μL antibody dilution buffer) and mixing thoroughly.

Avidin-Biotin-Peroxidase (ABC)
Diluted ABC solution has to be prepared freshly. Calculate the total volume needed for the assay by multiplying 0.1 mL/well and the number of wells required. Add 2-3 extra wells to the calculated number of wells to account for possible pipetting errors. Generate the required volume of diluted ABC solution by performing a 1:100 dilution (For each 1 μL concentrated ABC solution, add 99 μL ABC dilution buffer) and mixing thoroughly.

• Direct ELISA protocol
This is a general protocol in which antigen coating and blocking may not be required if the wells have been pre-adsorbed with the antigen by the manufacturer.

Antigen Coating
- Dilute purified antigens to a final concentration of 1-10 μg/ml in bicarbonate/carbonate antigen-coating buffer (100 mM NaHCO3 in deionized water; pH adjusted to 9.6). Pipette 100 μL of diluted antigen to each well of a microtiter plate.
- Cover the plate with adhesive plastic and incubate at 4°C overnight or 37°C for 30 minutes.
- Remove the coating solution and wash the plate 3 times with 200 μL PBS buffer with for 5 minutes each time. The coating/washing solutions can be removed by flicking the plate over a sink. The remaining drops can be removed by patting the plate on a paper towel or by aspiration. Do not allow the wells to dry out at any time.

Blocking
- Pipette 200 μL blocking buffer (5% w/v non-fat dry milk in PBS buffer) per well to block residual protein-binding sites. Alternatively, BSA or BlockACE can be used to replace non-fat dry milk.
- Cover the plate with adhesive plastic and incubate for 1-2 hour(s) at 37°C (or at 4°C overnight).
- Remove the blocking solution and wash the plate twice with 200 μL PBS for 5 min each time. Flick the plate and pat the plate as described in the coating step.

Reagent Preparation
Prepare for the diluted standard solutions as shown on above.

Primary Antibody Incubation
- Serially dilute the conjugated primary antibody with blocking buffer immediately before use. The optimal dilution should be determined by a titration assay according to the antibody manufacturer.
- Pipette 100 μL of diluted secondary antibody solution to each well.
- Cover the plate with adhesive plastic and incubate for 2 hours.
- Remove the content in the wells and wash them 3 times with 200 μL PBS buffer for 5 min each time. Flick and pat the plate as described in the coating step.

Substrate Preparation
Prepare the substrate solution immediately before use or bring the pre-made substrate to room temperature. The two widely used enzymes for signal detection are horse radish peroxidase (HRP) and alkaline phosphatase (AP), and their corresponding substrates, stopping solutions, detection absorbance wavelengths and colour developed are as follows:

EnzymeSubstrate*Stop SolutionAbsorbance (nm)Colour Developed
HRPTMB2M H2SO4450Yellow
ARPpNPP0.75M NaOH450Yellow
TMB: 3,3’,5,5’-tetramethylbenzidine; pNPP: p-nitrophenyl-phosphate

Note:
• The TMB substrate must be kept at 37°C for 30 min before use.
• Hydrogen peroxide can also act as a substrate for HRP. • Sodium azide is an inhibitor of HRP. Do not include the azide in buffers or wash solutions if HRP-labeled conjugate is used for detection.

Signal Detection
- Pipette 90 μL of substrate solution to the wells with the control and standard solutions.
- Incubate the plate at 37°C in the dark. If TMB is used, shades of blue will be observed in the wells with the most concentrated solutions. Other wells may show no obvious colour.
- Colour should be developed in positive wells after 15 min. After sufficient colour development, pipette 100 μL of stopping solution to the wells (if necessary).
- Read the absorbance (OD: Optical Density) of each well with a plate reader.

Data Analysis
- Prepare a standard curve using the data produced from the diluted standard solutions. Use absorbance on the Y-axis (linear) and concentration on the X-axis (log scale).
- Interpret the sample concentration from the standard curve.

 Indirect ELISA protocol
This is a general protocol in which antigen coating and blocking may not be required if the wells from the manufacturer have been pre-adsorbed with the antigen.

Antigen Coating
- Dilute the purified antigens to a final concentration of 1-10 μg/mL in bicarbonate/carbonate antigen-coating buffer (100 mM NaHCO3 in deionized water; pH adjusted to 9.6).
- Pipette 100 μL of diluted antigen to each well of the microtiter plate.

Blocking
- Pipette 200 μL blocking buffer (5% w/v non-fat dry milk in PBS buffer) per well to block residual protein-binding sites. Alternatively, BSA or BlockACE can be used to replace the non-fat dry milk.
- Cover the plate with adhesive plastic and incubate for 1-2 hour(s) at 37°C or at 4°C overnight.
- Remove the blocking solution and wash the plate twice with 200 μL PBS for 5 minutes each time. Flick and pat the plate as described in the coating step.

Reagent Preparation
- Prepare for the diluted standard solutions as shown on above.

Primary Antibody Incubation
- Serially dilute the primary antibody of choice with blocking buffer. The optimal dilution should be determined by a titration assay according to the antibody manufacturer.
- Pipette 100 μL of each diluted antibody per well. Repeat in duplicate or triplicate for accuracy. The negative control should be species- and isotype-matched, as well as non-specific immunoglobulin diluted in PBS buffer.
- Cover the plate with adhesive plastic and incubate for 1 hour at 37°C or 2 hours at room temperature. These incubation times should be sufficient to achieve a strong signal. However, if a weak signal is observed, perform incubation overnight at 4°C to obtain a stronger signal.
- Remove the diluted antibody solution and wash the wells 3 times with 200 μL PBS for 5 minutes each time. Flick the plate and pat the plate as described in the coating step.

Secondary Antibody Incubation
- Serially dilute the conjugated secondary antibody with blocking buffer immediately before use. The optimal dilution should be determined by a titration assay according to the antibody manufacturer.
- Pipette 100 μL of diluted secondary antibody solution to each well.
- Cover the plate with adhesive plastic and incubate for 2 hours at room temperature.
- Remove the content in the wells and wash them 3 times with 200 μL PBS buffer for 5 minutes each time. Flick and pat the plate as described in the coating step.

Substrate Preparation
Prepare the substrate solution immediately before use or bring the pre-made substrate to room temperature. The two widely used enzymes for signal detection are horse radish peroxidase (HRP) and alkaline phosphatase (AP), and their corresponding substrates, stopping solutions, detection absorbance wavelengths and colour developed are as follows:

EnzymeSubstrate*Stop SolutionAbsorbance (nm)Colour Developed
HRPTMB2M H2SO4450Yellow
APpNPP0.75M NaOH450Yellow
TMB: 3,3’,5,5’-tetramethylbenzidine; pNPP: p-nitrophenyl-phosphate

Note: 
• The TMB substrate must be kept at 37°C for 30 min before use.

• Hydrogen peroxide can also act as a substrate for HRP.

• Sodium azide is an inhibitor of HRP. Do not include the azide in buffers or wash solutions if HRP-labeled conjugate is used for detection.

Sandwich ELISA protocol
Capture Antibody Coating
- Dilute the capture antibody to a final concentration of 1-10 μg/mL in bicarbonate/carbonate antigen-coating buffer (100 mM NaHCO3 in deionized water; pH adjusted to 9.6).
- Pipette 100 μL of diluted antibody to each well of a microtiter plate.
- Cover the plate with adhesive plastic and incubate at 4°C overnight (or 37°C for 30 min).
- Remove the coating solution and wash the plate 3 times with 200 μL PBS with for 5 minutes each time. The coating/washing solutions can be removed by flicking the plate over a sink. The remaining drops can be removed by patting the plate on a paper towel or by aspiration. Do not allow the wells to dry out at any time.

Blocking
- Pipette 200 μL blocking buffer (5% w/v non-fat dry milk in PBS buffer; Alternatively, BSA or BlockACE can be used to replace non-fat dry milk.) per well to block residual protein-binding sites.
- Cover the plate with adhesive plastic and incubate for 1-2 hour(s) at 37°C (or at 4°C overnight).
- Remove the blocking solution and wash the plate twice with 200 μL PBS for 5 minutes each time. Flick and tap the plate as described in the coating step.

Reagent Preparation
Prepare biotinylated antibody and ABC solutions as described previously for the diluted standard solutions.

Sample (Antigen) Incubation
- Serially dilute the sample with blocking buffer immediately before use. The optimal dilution should be determined by a titration assay according to the antibody manufacturer.
- Pipette 100 μL of each of the diluted sample solutions and control to each empty well. Repeat in duplicate or triplicate for accuracy. The negative control should be species- and isotype-matched, as well as non-specific immunoglobulin diluted in PBS buffer.
- Cover the plate with adhesive plastic and incubate for 2 hours at room temperature.
- Remove the content in the wells and wash them 3 times with 200 μL PBS buffer for 5 minutes each. Flick the plate and pat the plate as described in the coating step.

Biotinylated Antibody Incubation
- Pipette 100 μL of diluted antibody to the control, standard solutions and diluted samples.
- Cover the plate with adhesive plastic and incubate for 1 hour at 37°C (or 2 hours at room temperature). These incubation times should be sufficient to receive a strong signal. However, if a weak signal is observed, perform incubation overnight at 4°C for a stronger signal.
- Remove the content in the wells and wash them 3 times with 200 μL PBS for 5 min each time. Flick the plate and pat the plate as described in the coating step.

ABC Incubation
- Pipette 100 μL of diluted ABC solution to the wells with control, standard solutions and diluted samples.
- Cover the plate with adhesive plastic and incubate for 30 minutes at 37°C.
- Remove the content in the wells and wash them 3 times with 200 μL PBS buffer for 5 minutes each. Flick the plate and pat the plate as described in the coating step.

Substrate Preparation
Prepare the substrate solution immediately before use or bring the pre-made substrate to room temperature. The two widely used enzymes for signal detection are horse radish peroxidase (HRP) and alkaline phosphatase (AP), and their corresponding substrates, stopping solutions, detection absorbance wavelengths and colour developed are as follows:

EnzymeSubstrate*Stop SolutionAbsorbance (nm)Colour Developed
HRPTMB2M H2SO4450Yellow
APpNPP0.75M NaOH450Yellow
TMB: 3,3’,5,5’-tetramethylbenzidine; pNPP: p-nitrophenyl-phosphate

Note:
• The TMB substrate must be kept at 37°C for 30 minutes before use.
• Hydrogen peroxide can also act as a substrate for HRP.
• Sodium azide is an inhibitor of HRP. Do not include the azide in buffers or wash solutions if HRP-labeled conjugate is used for detection.

Signal Detection
- Pipette 90 μL of substrate solution to the wells with the control, standard solutions and diluted samples.
- Incubate the plate at 37°C in the dark. If TMB is used, shades of blue will be observed in the wells with the most concentrated solutions. Other wells may show no obvious colour.
- A colourimetric signal should be developed in positive wells after 15 minutes. Stop the reaction by adding 100 μL of stop solution.
- Read the absorbance (OD: Optical Density) of each well with a plate reader.

Data Analysis
- Prepare a standard curve using the data produced from the diluted standard solutions. Use absorbance on the Y-axis (linear) and concentration on the X-axis (log scale). Competitive ELISA yields an inverse curve: Higher values of antigen in the samples yield a smaller amount of colour change.
- Interpret the sample concentration from the standard curve.

Competitive ELISA protocol
This is a general protocol in which antigen coating and blocking may not be required if the wells have been pre-adsorbed with the antigen by the manufacturer.

Antigen Coating
- Dilute purified antigens to a final concentration of 20 μg/ml in bicarbonate/carbonate antigen-coating buffer (100 mM NaHCO3 in deionized water; pH adjusted to 9.6).
- Pipette 100 μL of diluted antigen to each well of a microtiter plate.
- Cover the plate with adhesive plastic and incubate at 4°C overnight or 37°C for 30 minutes.
- Remove the coating solution and wash the plate 3X with 200 μL PBS buffer with for 5 minutes each time. The coating/washing solutions can be removed by flicking the plate over a sink. The remaining drops can be removed by patting the plate on a paper towel or by aspiration. Do not allow the wells to dry out at any time.

Blocking
- Pipette 200 μL blocking buffer (5% w/v non-fat dry milk in PBS buffer) per well to block residual protein-binding sites. Alternatively, BSA or BlockACE can be used to replace non-fat dry milk.
- Cover the plate with adhesive plastic and incubate for 1-2 hour(s) at 37°C or at 4°C overnight.
- Remove the blocking solution and wash the plate twice with 200 μL PBS for 5 minutes each time. Flick and pat the plate as described in the coating step.

Reagent Preparation - Prepare for the diluted standard solutions as shown on p.9.

Sample Incubation
- Serially dilute the sample with blocking buffer immediately before use. The optimal dilution should be determined by a titration assay according to the antibody manufacturer.
- Pipette 100 μL of diluted sample to each well.
- Cover the plate with adhesive plastic and incubate for 2 hours at room temperature.
- Remove the content in the wells and wash them three times with 200 μL PBS buffer for 5 minutes each time. Flick and pat the plate as described in the coating step.

Primary Antibody Incubation
- Serially dilute the primary antibody of choice with blocking buffer. The optimal dilution should be determined by a titration assay according to the antibody manufacturer.
- Pipette 100 μL of each diluted antibody per well. Repeat in duplicate or triplicate for accuracy. The negative control should be species- and isotype-matched as well as non-specific immunoglobulin diluted in PBS buffer.
- Cover the plate with adhesive plastic and incubate for 1 hour at 37°C or 2 hours at room temperature. These incubation times should be sufficient to receive a strong signal. However, if a weak signal is observed, perform incubation overnight at 4°C for a stronger signal.
- Remove the diluted antibody solution and wash the wells 3 times with 200 μL PBS for 5 minutes each time. Flick the plate and pat the plate as described in the coating step.

Secondary Antibody Incubation
- Serially dilute the conjugated secondary antibody with blocking buffer immediately before use. The optimal dilution should be determined by a titration assay according to the antibody manufacturer.
- Pipette 100 μL of diluted secondary antibody solution to each well.
- Cover the plate with adhesive plastic and incubate for 2 hours at room temperature.
- Remove the content in the wells and wash them 3 times with 200 μL PBS buffer for 5 minutes each time. Flick the plate and pat the plate as described in the coating step.

Substrate Preparation
Prepare the substrate solution immediately before use or bring the pre-made substrate to room temperature. The two widely used enzymes for signal detection are horse radish peroxidase (HRP) and alkaline phosphatase (AP), and their corresponding substrates, stopping solutions, detection absorbance wavelengths and colour developed are as follows:

EnzymeSubstrate*Stop SolutionAbsorbance (nm)Colour Developed
HRPTMB2M H2SO4450Yellow
APpNPP0.75M NaOH450Yellow
TMB: 3,3’,5,5’-tetramethylbenzidine; pNPP: p-nitrophenyl-phosphate

Note:
• The TMB substrate must be kept at 37°C for 30 minutes before use.
• Hydrogen peroxide can also act as a substrate for HRP.
• Sodium azide is an inhibitor of HRP. Do not include the azide in buffers or wash solutions if HRP-labeled conjugate is used for detection.

Signal Detection
- Pipette 90 μL of substrate solution to the wells with the control, standard solutions and diluted samples.
- Incubate the plate at 37°C in the dark. If TMB is used, shades of blue will be observed in the wells with the most concentrated solutions. Other wells may show no obvious colour.
- Colour should be developed in positive wells after 15 minutes. After sufficient colour development, pipette 100 μL of stopping solution to the wells (if necessary).
- Read the absorbance (OD: Optical Density) of each well with a plate reader.

Data Analysis
- Prepare a standard curve using the data produced from the diluted standard solutions. Use absorbance on the Y-axis (linear) and concentration on the X-axis (log scale).
- Competitive ELISA yields an inverse curve: Higher values of antigen in the samples yield a smaller amount of colour change.
- Interpret the sample concentration from the standard curve

7. Troubleshooting Guide

The following guide serves as a checklist for the possible causes and solutions with respect to some of the most commonly encountered problems from the ELISA assays.

• Weak / no signal

 Possible causeSolution to the problem
1Blocking protein in coating solutionEliminate blocking protein from coating solution
2Capture antibody (or antigen) does not bind to plateUse ELISA plate, not tissue culture plate Extend coating time Increase concentration of coating components
3Problem with the standardUse new sample Check that the standard is appropriately handled
4Shortened incubation timeFollow the manufacturer guideline (If the problem persists, try incubating samples at 4°C overnight)
5Incubation temperature too lowEnsure incubations are done at correct temperature Before proceeding, all reagents, including plate, should be at room temperature or as recommended by the manufacturer
6Incompatible sample typeUse sample that the assay is known to detect a positive control (Include such control in your experiment)
7Incompatible assay bufferEnsure assay buffer is compatible with the target of interest
8Target present below detection limitDecrease dilution factor or concentrate samples
9Incorrect/Insufficient/No substrateCheck the substrate identity Increase concentration or amount of substrate Follow manufacturer guidelines
10Incorrect/Insufficient/No antibodyCheck the antibody identity Repeat the assay with higher antibody concentrations to find the optimal one for your experiment
11Antibody stored at 4°C for several weeks or subjected to repeated freeze-thaw cyclesUse fresh aliquot of antibody that has been stored at -20°C or below
12Incorrect reagents added/ prepared; Missing reagentsCheck protocol, ensure correct reagents are added in proper order and prepared to correct concentrations (e.g. TMB for HRP-labeled antibodies)
13Expired/Contaminated reagentsMake and use fresh/uncontaminated reagents
14Enzyme inhibitor presentAvoid sodium azide in HRP reactions Avoid phosphate in AP reactions
15Incorrect storage of componentsDouble check storage conditions on kit level (Most kits need to be stored at 4°C)
16Ultra-vigorous plate washingGently pipette wash buffer (manual method) Ensure correct pressure (automatic wash system)
17Dried out wellsCover the plate using sealing film or tape for all incubations
18Wells scratched with pipette or pipette tipsCarefully dispense/aspirate solutions into and out of wells
19Plate read at incorrect detection wavelengthUse recommended wavelength/filter Ensure plate reader is set correctly for type of substrate used
20Slow colour developmentPrepare substrate immediately before use Allow longer incubation Ensure stock solution is unexpired and uncontaminated
21Epitope recognition impeded by adsorption to plateConjugate peptide to large carrier protein before coating onto plate

• Saturated signal

 Possible causeSolution to the problem
1High sample concentrationUse higher sample dilutions (Determine the optimal dilutions by titration assay)
2Excessive substrateDecrease concentration or amount of substrate: Follow manufacturer guidelines (The substrate provided with the ELISA kit might require further dilution)
3Substrate colour changed beforeMake substrate immediately before use
4Unspecific antibody bindingEnsure wells are pre-processed to prevent non-specific binding.
Use affinity-purified antibody and preferably one that is pre-adsorbed.
Use serum (5-10%) from same species as secondary antibody (bovine serum is also recommended)
5Exceeded incubation timeFollow the manufacturer guidelines (If the problem persists, try incubating samples at 4°C overnight)
6Excess antibodyRepeat the assay with lower antibody concentrations to find the optimal one for your experiment
7Contaminated buffers with metals or HRPMake and use fresh buffers
8Insufficient washingFollow the manufacturer guidelines At the end of each washing step, flick the plate over a sink and pat the plate on a paper towel
9Plate sealers not used or re-usedUse recommended wavelength/filter.
Ensure plate reader is set correctly for type of substrate used.
10Plate read at incorrect detection wavelengthConjugate peptide to large carrier protein before coating onto plate
11Excess time before plate readingRead your plate within 30 minutes after adding the substrate (If the reading is not performed within this time frame, add a stopping solution after sufficient colour is developed in the plate)

• High Background

 Possible causeSolution to the problem
1Insufficient washingFollow the manufacturer guidelines
At the end of each washing step, flick the plate over a sink and pat the plate on a paper towel
2Ineffective/Contaminated blocking bufferTry higher blocking protein concentration
Increase blocking time
Use fresh buffer
3Excess antibodyRepeat the assay with lower antibody concentrations to find the optimal one for your experiment
4Excess substrateDecrease concentration or amount of substrate Follow manufacturer guidelines (Note: The substrate provided with the ELISA kit might require further dilution)
5Cross reactivity (Detection antibody reacts with coating antibody)Run appropriate controls
6Unspecific antibody bindingTry different formulations in coating solutions
Ensure wells are pre-processed to prevent non-specific binding
Use affinity-purified antibody and preferably one that is pre-adsorbed
Use serum (5-10%) from same species as secondary antibody (bovine serum is also recommended)
7Insufficient Tween in buffersUse PBS containing 0.05% Tween
8Suboptimal salt concentration in washing bufferOptimize salt concentration as high concentration can reduce non-specific interactions
9Exceeded incubation temperatureOptimize incubation temperature for your assay (antibodies bind optimally at very specific temperature)
10Insufficient reagent mixtureThoroughly mix all reagents and samples before pipetting solutions into wells
11Blanks contaminated with samplesChange pipette tips when switching between blanks and samples
Put a lid on plates to avoid any spilling between wells
12Sample contaminated with enzymesTest samples with substrate alone to check for contaminating enzymes
13Contaminated TMB substrateUse a clean container to check that the substrate is not contaminated (TMB substrate should be clear and colourless before adding to wells)
14Substrate incubation in lightCarry out substrate incubation in dark or follow recommendation from manufacturer
15Uneven evaporation of solution from wells during incubationAlways incubate with a lid on the plate
16Precipitate created in wells upon substrate additionIncrease dilution factor of sample or decrease concentration of substrate
17Exceeded incubation timeFollow the manufacturer guidelines (If the problem persists, try incubating samples at 4°C overnight)
18Incorrect standard curve dilutionsCheck pipetting techniques.
Double check calculations.
19Plates stacked during incubations, leading to uneven temperature distributionAvoid stacking plates
20Dirty or defective platesClean the plate bottom
21Unstopped colour developmentUse Stopping solution to prevent over-development
22Excess time before plate readingRead your plate within 30 minutes after adding the substrate (If the reading is not performed within this time frame, add a stopping solution after sufficient colour is developed in the plate)
Note: Colour continues to develop even after adding the stopping solution (although at a slower rate)
23Incorrect plate reading settingUse recommended wavelength/filter
Ensure plate reader is set correctly for type of substrate used

• Low sensitivity

 Possible causeSolution to the problem
1Assay format not sensitive enoughSwitch to a more sensitive detection system (e.g. colourimetric to chemiluminescence)
Switch to a more sensitive assay type (e.g. direct ELISA to sandwich ELISA)
Increase incubation time and/or temperature
2Improper storage of ELISA kitStore all reagents as recommended (Note: all reagents may not have identical storage requirements)
3Insufficient targetReduce sample dilution or concentrate sample
4Inactive substrateEnsure reporter enzyme has the expected activity
5Poor target adsorption to wellsCovalently link target to wells
6Insufficient substrateIncrease concentration or amount of substrate
7Incompatible sample typeUse a sample that the assay is known to detect a positive control.
Include positive control in your experiment.
8Interfering ingredients in buffers and sampleCheck reagents for any interfering chemicals, e.g. sodium azide in antibodies inhibit HRP enzyme; EDTA used as anti-coagulant for plasma collection inhibits enzymatic reactions
9Mixing or substituting reagents from different kitsAvoid mixing components from different kits
10Incorrect plate reading settingUse recommended wavelength/filter Ensure plate reader is set correctly for type of substrate used

• Poor standard curve

 Possible causeSolution to the problem
1Improper standard solutionConfirm dilutions are done correctly.
Make new standard curve as appropriate.
2Standard improperly reconstitutedBriefly spin vial before opening.
Inspect for undissolved material after reconstituting.
3Standard degradedStore and handle standard as recommended.
Prepare standards no more than two hours before use.
4Improper curve fittingTry plotting using different scales, e.g. log-log, 5-parameter logistic curve fit
5Pipetting errorUse calibrated pipettes and proper pipetting technique
6Insufficient washingFollow the manufacturer guidelines. At the end of each washing step, flick the plate over a sink and pat the plate on a paper towel
7Insufficiently mixed reagentsThoroughly mix reagents
8Poor/variable adsorption of reagents to plateExtend incubation time.
Check coating buffer.
Use a different plate as appropriate.
Check homogeneity of samples.
9Plates stacked during incubationKeep plates separated if not using rotating plates
10Dirty or defective platesClean the plate bottom

• Poor replicate data

 Possible causeSolution to the problem
1Bubble in wellsEnsure no bubbles are present prior to reading plate
2Insufficient washing of wellsCarefully wash wells.
Follow recommended protocols.
Check that all ports of the plate washer are unobstructed.
3Incomplete reagent mixingEnsure all reagents are mixed thoroughly
4Inconsistent pipettingUse calibrated pipettes and proper pipetting techniques.
If a multi-channel pipette is used, ensure that all channels deliver the same volume.
5Inconsistent sample prep or storageEnsure consistent sample prep and optimal sample storage conditions (e.g. minimize freeze/thaw cycles)
6Particulates in samplesRemove the particulates by centrifugation
7Plate sealers not used or re-usedDuring incubations, cover plates with plate sealers.
Use a fresh sealer every time the used sealer is removed from the plate.
8Cross-well contaminationEnsure plate sealers and pipette tips are not contaminated with reagents
9Edge effect (higher or lower OD in peripheral wells than in central wells)Ensure plates and reagents are kept at room temperature before pipetting into wells unless otherwise instructed.
During incubation, seal the plate completely with a plate sealer and avoid stacking plates.

• Inconsistent Assay-to-Assay results

 Possible causeSolution to the problem
1Insufficient washing of wellsCarefully wash wells.
Follow recommended protocols.
Check that all ports of the plate washer are unobstructed.
2Variation in incubation temperatureAdhere to recommended incubation temperature.
Avoid incubating plates in area where environmental conditions vary.
3Variation in protocolAdhere to the same protocol from run to run
4Plate sealers not used or re-usedDuring incubations, cover plates with plate sealers.
Use a fresh sealer every time the used sealer is removed from the plate.
5Incorrect dilutionsConfirm dilutions are done correctly for standard solutions, etc.
Make new standard curve as appropriate.
6Contaminated buffersMake and use fresh buffers
7Plates stacked during incubationKeep plates separated if not using rotating plates

• Slow colour development

 Possible causeSolution to the problem
1Substrates too old, contaminated or used at incorrect pHMake and use fresh substrates at correct pH: they should be prepared immediately before use
2Expired/Contaminated solutionsMake and use fresh reagents
3Incorrect incubation temperatureEnsure plates and reagents are kept at room temperature before pipetting into wells unless otherwise instructed.
During incubation, seal the plate completely with a plate sealer and avoid stacking plates.
4Low antibody concentrationRepeat the assay with higher antibody concentrations to find the optimal one for your experiment
5Low substrate concentrationAdd more substrate to the wells.
Make substrate no more than one hour before use.
Note: Typical ELISA sensitivity is ~0.1 pg/mL with exact value depends on antibody used.

• Plate imaging problems

 Possible causeSolution to the problem
1Oversaturated image after acquisitionUse full resolution image to analyze results (Do not use jpeg or other compressed formats)
2Blurry spots in imagesRe-focus your camera before taking a new image
3Repeated pixel values or rectangular spotsUse lower bin size, higher image resolution and/or lossless file type sealer and avoid stacking plates
4Flat standard in imagesReduce acquisition time

Gain Data

 Arigo is proud to launch an in-house developed GainDataTM (ELISA analysis tool, web based free service).

Researchers just simply copy their ELISA data from excel and paste to our GainDataTM  then  follow the tool, less than a minute the result will be calculated automatically.